CALIBRATION OF pH METER (Make: Thermo Orion)

  • OBJECTIVE:
  • To lay down the procedure for calibration of pH meter (Make: Thermo Orion)
  • SCOPE:

This SOP is applicable to the procedure for calibration of pH meter (Make: Thermo Orion) at {Company Name} {Location}.

  • RESPONSIBILITY:
  • Officer/Executive/Designee Quality Control – Shall be responsible for operation as per SOP.
  • Head/Designee Quality Control – Shall be responsible for ensuring compliance as per SOP.
  • ACCOUNTABILITY:

             QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:
    • Press POWER key and operate the instrument as per the SOP.
    • Press Cal key to begin calibration.
    • Place the electrode and ATC probe into the first buffer which has transferred from mother container to cleaned dried beaker. The neutral buffer pH 7.00 is usually recommended to be used first followed from lowest to highest pH.
    • The display shows the reading of CAL 1 (7.00 pH)
    • Press CAL key after the reading of CAL 1 becomes ready.
    • Place the electrode in 2nd buffer.
    • The display then shows Cal 2 (4.01 pH).
    • Press Cal key after the reading of CAL 2 becomes ready.
    • Place the electrode in 3rd buffer.
    • The display then shows Cal 3 (10.01)
    • Press measure key to end calibration.
    • The slope will be displayed and it should be between 92-102.
    • Buffers should be disposed of after calibration.

Note: To avoid Contamination, transfer the required amount of buffer into a clean dry beaker and discard the solution after Calibration is completed.

  • In case of unavailability of commercial buffers prepare the buffers as per given Format–II and calibration readings of the same shall be record in Format-III.
  • REFERENCES:

Not Applicable

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IpH Meter calibration record.
Annexure-IIpH Buffer preparation record
Annexure-IIIpH Meter calibration record for manual Buffers

         ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
SOP:Standard Operating Procedure
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

               Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Annexure-I
pH METER CALIBRATION RECORD

Annexure-II
pH BUFFER PREPARATION RECORD

Annexure-III
pH METER CALIBRATION RECORD FOR MANUAL BUFFERS

Share this article
Shareable URL
Prev Post

ANALYTICAL WORK SHEET AND COA PREPARATION

Next Post

BRACKETING OF STANDARDS

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading