- OBJECTIVE:
- To lay down the procedure for calibration of pH meter (Make: Thermo Orion)
- SCOPE:
This SOP is applicable to the procedure for calibration of pH meter (Make: Thermo Orion) at {Company Name} {Location}.
- RESPONSIBILITY:
- Officer/Executive/Designee Quality Control – Shall be responsible for operation as per SOP.
- Head/Designee Quality Control – Shall be responsible for ensuring compliance as per SOP.
- ACCOUNTABILITY:
QA Head shall be Accountable for implementation of SOP.
- PROCEDURE:
- Press POWER key and operate the instrument as per the SOP.
- Press Cal key to begin calibration.
- Place the electrode and ATC probe into the first buffer which has transferred from mother container to cleaned dried beaker. The neutral buffer pH 7.00 is usually recommended to be used first followed from lowest to highest pH.
- The display shows the reading of CAL 1 (7.00 pH)
- Press CAL key after the reading of CAL 1 becomes ready.
- Place the electrode in 2nd buffer.
- The display then shows Cal 2 (4.01 pH).
- Press Cal key after the reading of CAL 2 becomes ready.
- Place the electrode in 3rd buffer.
- The display then shows Cal 3 (10.01)
- Press measure key to end calibration.
- The slope will be displayed and it should be between 92-102.
- Buffers should be disposed of after calibration.
Note: To avoid Contamination, transfer the required amount of buffer into a clean dry beaker and discard the solution after Calibration is completed.
- In case of unavailability of commercial buffers prepare the buffers as per given Format–II and calibration readings of the same shall be record in Format-III.
- REFERENCES:
Not Applicable
- ANNEXURES:
ANNEXURE NO. | TITLE OF ANNEXURE |
Annexure-I | pH Meter calibration record. |
Annexure-II | pH Buffer preparation record |
Annexure-III | pH Meter calibration record for manual Buffers |
ENCLOSURES: SOP Training Record.
- DISTRIBUTION:
- Controlled Copy No. 01 : Head Quality Assurance
- Controlled Copy No. 02 : Head Quality Control
- Master Copy : Quality Assurance Department
- ABBREVIATIONS:
No. | : | Number |
SOP | : | Standard Operating Procedure |
QC | : | Quality Control |
- REVISION HISTORY:
CHANGE HISTORY LOG
Revision No. | Details of Changes | Reason for Change | Effective Date |
00 | New SOP | Not Applicable | To Be Written Manual |
Annexure-I
pH METER CALIBRATION RECORD
Annexure-II
pH BUFFER PREPARATION RECORD
Annexure-III
pH METER CALIBRATION RECORD FOR MANUAL BUFFERS