CALIBRATION OF TAP DENSITY APPARATUS

  • OBJECTIVE:
  • To lay down the procedure for Calibration of Tap density apparatus.
  • SCOPE:
  • This SOP is applicable to the procedure for Calibration of Tap density apparatus at {Company Name} {Location}.
  • RESPONSIBILITY:
  • Executive/Designee Quality Control is responsible to calibrate the instrument as per SOP.
  • Head QC: Ensure the compliance of SOP.
  • ACCOUNTABILITY:

             QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:
    • Before switching on the instrument, check the area and instrument cleanliness.
    • Connect the plug to the power supply and switch on the power.
    • Operate the instrument as per SOP.
    • The following parameters to be calibrated for Tap density apparatus:
  • Drop height of cylinder holder by using digital vernier caliper
  • Number of drops per minute by using stopwatch
    • Calibration of Drop height of cylinder holder for USP method 1:
      • Open the top cover by removing bottom screws.
      • Turn the cam attached to the motor shaft clock wise, so that the cylinder holder shaft is at the minimum position.
      • Make sure that the bottom plate of the cylinder holder is in perfect contact with the tapping platform.
      • Measure the distance between the top plate of frame and the clamp plate of cylinder holder by using digital vernier caliper; this reading is the base height (HR) as shown in figure (a).
      • Rotate the cam further by hand, clockwise till the cylinder holder shaft reaches maximum height.
      • At this position the shaft tip should be on the falling edge of the cam arm. Mark this arm No; 1 as shown in figure (b).
      • Measure the distance between the top plate of frame and clamp plate of cylinder holder by using digital vernier caliper. This is height HI.
  • Make a mark on the point of measurement on the top plate of frame and on the cylinder plate.
    • The difference between these two readings is drop height of the cylinder.
    • Mark arms 1, 2 and 3 and heights H1, H2, H3 and so on. Repeat the above procedure for all arms with both cylinder holders.
    • Make sure that the point of measurement on the top plate of frame and on the cylinder holder plate is maintained the same for all measurements.
    • Record the results in the calibration format as per Format-I.

Click the link for download word file copy of this document:
https://pharmaguidehub.com/product/calibration-of-tap-density-apparatus/

  • Repeat the same procedure for all arms with both cylinder holder (figure c) and record the results in the calibration format.
    • Calibration for Number of drops of cylinder holder per minute:
      • Choose the mode menu to ‘USP1’ by using digital scroll key.
      • Press set key, set taps count to 300.
      • Press start key to start the tapping and start the stop watch.
      • Stop the stop watch after 1 Minute and note down the number of drops.
      • The number of taps should be 300 ± 6.
      • Similarly choose the mode menu to ‘USP11’ by using digital scroll key.
      • Press set key, set taps count to 250.
      • Press start key to start the tapping and start the stop watch.
      • Stop the stop watch after 1 Minute and note down the number of drops.
      • The number of drops should be 250 ± 5.
  • REFERENCES:

Not Applicable

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-ITap density apparatus Calibration Record

        ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
SOP:Standard Operating Procedure
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

               Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Annexure-I
TAP DENSITY APPARATUS CALIBRATION RECORD

Click the link for download word file copy of this document:
https://pharmaguidehub.com/product/calibration-of-tap-density-apparatus/

Share this article
Shareable URL
Prev Post

CALIBRATION OF REFRACTOMETER (MAKE – RUDOLPH RESEARCH)

Next Post

TESTING AND RELEASE OF RAW MATERIALS

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

HANDLING OF DESICCATOR

To lay down the procedure for handling and Maintenance of Desiccator. This SOP is applicable for the procedure…

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading