- OBJECTIVE:
To lay down the procedure for Validation of different types of analytical test procedures.
- SCOPE:
This SOP is applicable for the procedure for Validation of different types of analytical test procedures at {Company Name} {Location}.
- RESPONSIBILITY:
- Officer/Executive/Designee Quality Control – Shall be responsible for conducting validation of analytical test procedures as per SOP.
- Head/Designee Quality Control – Shall be responsible for ensuring compliance as per SOP.
- ACCOUNTABILITY:
QA Head shall be Accountable for implementation of SOP.
- PROCEDURE:
- All analytical methods shall be validated in accordance with current ICH Q1 requirements.
- According to the current good manufacturing Practice Regulations [ (21CFR 211.194) (a) (2)], users of analytical methods described in (Compendial Method) are not required to validate accuracy and reliability of these methods, but merely to verify their suitability under actual conditions of use.
- Analytical methods used for evaluation of cleaning process validation shall be done at Quality Control in accordance to the above guidelines.
- Cleaning Validation Procedure:
- Specificity:
- Specificity is the ability to assess unequivocally the analyte in the presence of components which may be expected to be present.
- Demonstrate the separation of the analyte from available placebo, Diluent, blank(swab) and placebo with standard solution using the Instrument.
- System Precision:
- The precision of an analytical procedure expresses the closeness of the agreement (Degree of scatter) between a series of measurements.
- As a part of method validation, a minimum of 6 injections of the standard preparation with an RSD of ≤ 5.0 % is recommended.
- Method Precision:
- The precision of an analytical procedure expresses the closeness of the agreement (Degree of scatter) between a series of measurements obtained from a multiple sampling of the same homogeneous sample under the prescribed conditions.
- As a part of method validation, a minimum of 6 injections of the standard preparation with an RSD of ≤ 5.0 % is recommended.
- Linearity:
- The linearity of an analytical procedure is its ability (with a given range) to obtain test results which are directly proportional to the concentration (amount) of analyte in the sample.
- The Correlation coefficient should be greater than 0.99.
- Limit of detection:
- The limit of detection of an individual analytical procedure is the lowest amount of the analyte in a sample which can be detected but not necessarily quantified as an exact value.
- Calculate LOD by using (3.3 x Steyx) / slope. Prepare and inject LOD solution accordingly. The % RSD should not be more than 33.0%.
- Limit of Quantification:
- The Quantification limit of an individual analytical procedure is the lowest amount of the analyte in a sample which can be quantitatively determined with a suitable precision and accuracy. The Quantification limit is a parameter of quantitative assays for low levels of compounds in sample matrices and is used particularly for the determination of impurities and/or degradation products.
- Calculate LOQ by using (10 x Steyx) / slope. Prepare and inject LOQ solution (accordingly. The % RSD should not be more than 10.0 %.
- Percentage Recovery:
- Standard Solutions: Prepare standard solutions corresponding to approximately 50%, 100% and 150% of Acceptance Limit concentration as per the test procedure.
- Spiked Swab solution: Prepare spiked swab solutions corresponding to approximately 50%, 100% and 150% of Acceptance Limit as per the test procedure.
- Calculate the % Recovery of spiked swab using the following formula.
- Specificity:
% Recovery = (Spiked swab Area / Average Standard Area) x 100
- Average % Recovery should be not less than 75%. In case of discrepancy repeat the experiment. If these criteria are still not met, investigate the reason.
- Numbering of ACMV protocol and report:
- Analytical Cleaning Method validation Protocol number consists of 15 characters.
- Numbering of ACMV protocol and report:
ACMVP/QC/YY-XXX
where
ACMVP | : | Analytical Cleaning Method Validation protocol |
/ | : | slash |
QC | : | Quality control |
/ | : | slash |
YY | : | Current year |
– | : | Dash |
XXX | : | three digits serial number starts from 001,002 and so on |
E.g.: First Analytical Cleaning Method Validation protocol of 2024 in QC shall be given as ACMVP/QC/24-001
- Analytical Cleaning Method validation Report numbering system consist of 15 characters
ACMVR/QC/YY-XXX
where
ACMVR | : | Analytical Cleaning Method Validation report |
/ | : | slash |
QC | : | Quality control |
/ | : | slash |
YY | : | Current year |
– | : | Dash |
XXX | : | three digits serial number starts from 001,002 and so on |
E.g.: First Analytical Cleaning Method Validation report of 2024 in QC shall be given as ACMVR/QC/24-001.
- The Analytical Cleaning Method validation Protocol and Report along with supporting data shall be submitted to QA.
- REFERENCES:
Not Applicable
- ANNEXURES:
ANNEXURE NO. | TITLE OF ANNEXURE |
Annexure-I | Analytical Cleaning Method validation Record |
ENCLOSURES: SOP Training Record.
- DISTRIBUTION:
- Controlled Copy No. 01 : Head Quality Assurance
- Controlled Copy No. 02 : Head Quality Control
- Master Copy : Quality Assurance Department
- ABBREVIATIONS:
No. | : | Number |
HPLC | : | High Performance Liquid Chromatography |
RSD | : | Relative standard deviation |
SOP | : | Standard Operating Procedure |
QC | : | Quality Control |
- REVISION HISTORY:
CHANGE HISTORY LOG
Revision No. | Details of Changes | Reason for Change | Effective Date |
00 | New SOP | Not Applicable | To Be Written Manual |
Annexure-I
ANALYTICAL CLEANING METHOD VALIDATION LOGBOOK