CALIBRATION OF VACUUM OVEN MAKE: NEWTRONIC

  • OBJECTIVE:
  • To lay down the procedure for Calibration of Vacuum Oven, Make: Newtronic.
  • SCOPE:

This SOP is applicable to the procedure for Calibration of Vacuum Oven at {Company Name} {Location}.

  • RESPONSIBILITY:
  • Executive/Designee Quality Control is responsible to calibrate the instrument as per SOP.
  • Head QC: Ensure the compliance of SOP.
  • ACCOUNTABILITY:

             QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:

A vacuum oven is a specialized oven that removes moisture and other gases from a substance by creating a low-pressure environment inside the chamber. This is different from a standard oven, which uses heat and air circulation to dry materials.

  • Operate the instrument as per SOP.
    • Set the temperature controller knob to 60°C.
    • After the set temperature is attained, measure it with a calibrated thermometer.
    • The observed temperature should be within ± 2.0°C.
    • Similarly perform the test at 100°C and 150°C and measure the temperature with a calibrated thermometer.
    • The observed temperature should be within ± 2.0°C.
    • Record the readings in vacuum oven calibration record as per Format-I.
    • Create a vacuum of 400 mm and close the valve and observe the Vacuum drop for 3 hours. The Vacuum drop should be not more than 1.0% of the initial Pressure after 3 hours.
    • The frequency for calibration of vacuum oven is Six months.
    • If the calibration is not proper, report the results to the departmental head for taking an appropriate action.

Click the link for download word file copy of this document:
https://pharmaguidehub.com/product/calibration-of-vacuum-oven-make-newtronic/

  • REFERENCES:

Not Applicable

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IVacuum Oven Calibration record

        ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
SOP:Standard Operating Procedure
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Annexure-I
VACUUM OVEN CALIBRATION RECORD

Click the link for download word file copy of this document:
https://pharmaguidehub.com/product/calibration-of-vacuum-oven-make-newtronic/

Share this article
Shareable URL
Prev Post

CALIBRATION OF MELTING POINT APPARATUS

Next Post

CALIBRATION OF ALLIANCE HPLC WITH FLUORESCENCE DETECTOR

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

OPERATION OF UV CABINET

QA Head shall be Accountable for implementation of SOP. Click the link to download word file copy of this…

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading