Your Compass for Pharma Success

CALIBRATION OF AUTOTITRATOR AND ELECTRODES

  • OBJECTIVE:

To lay down the procedure for Calibration of auto titrator and electrodes.

  • SCOPE:
  • This SOP is applicable to the procedure for Calibration of auto titrator and electrodes at {Company Name} {Location}.
  • RESPONSIBILITY:
  • Officer/Executive/Designee Quality Control – Shall be responsible for calibration of instrument as per SOP.
  • Head/Designee Quality Control – Shall be responsible for ensuring compliance as per SOP.
  • ACCOUNTABILITY:

QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:
    • Calibration of burette:
      • Operate the instrument as per Sop, Rinse well the burette and fill the Burette with purified water.Take a clean, dry 10ml volumetric flask.Weigh the volumetric flask and note down the weight as W1.Set the dispense volume as 1 ml and dose the water into the Volumetric flask.Weigh the volumetric flask with water and note weight as W2.Repeat the above 3 steps for volumes 2,3,4,5,6,7,8,9,10 respectively.Mread = W2 – W1
      • Calculate the actually dispensed water in a volumetric flask using the formula:
  • Vactual = Mread x Density
  • Calculate the relative error using the formula
  • Relative error = (Vactual –Vset) x 100/Vset
    • Density calculation:
      • Take a pycnometer and note down the empty weight (W1).
      • Fill the pycnometer with water, close the stopper provided and wipe off the excess water.
      • Now note down the weight of pycnometer with water at 25°C (W2).
      • Weight per ml of water = (W2-W1)/Volume of Pycnometer.
      • Enter the details of calibration in calibration record as per format-I.
      • Acceptance criteria: the average of Relative error should not be more than ± 0.5%
    • Auto titrator calibration:
      • Select a standard titration (KHP versus NaOH) suggested by Metrohm.
      • Prepare 0.1M NaOH solution and perform the standard titration for five times.
      • Calculate the molarity of NaOH for five titrations.
      • Calculate the RSD for molarity using the formula RSD = Standard Deviation % 100 Mean
      • Acceptance criteria: RSD value should not be more than ± 0.5%.
      • Enter the details in calibration record as per format-II.
      • If the calibration is not within the specified limits repeat the calibration. Report the Results to the department head for appropriate action.
    • Electrodes Calibration:
      • pH Electrodes:
        • Check the pH with three buffers 7.00, 4.01 and 10.01 and record the results in Format-III.
        • Acceptance criteria: The slope value should lie between 95% and 105% and Individual pH should lie within 0.05.
      • Solvotrode:
        • Check the mV value for buffers pH 4.0 and pH 7.0 and record the results in Format-IV.
        • Acceptance criteria: The difference between the two mV values should be Not less than150 mV.
      • Combined Ag ring Electrode:
        • Prepare a 0.1 N AgNO3 titrant and standardize using NaCl primary standard and check the purity of NaCl by carrying an assay titration and record the Results in Format-V.
        • Standardize and perform the titration as per GTP.
        • Acceptance criteria: The purity should lie within 99% to 101%.
      • Double Platinum Electrode Serial Number:
        • This is an electrode with two platinum wires so no calibration is required.
        • It is enough to ensure the two platinum wires are parallel to each other and clean it with a soft tissue dipped in acetone.
  • REFERENCES:

Not Applicable

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IBurette calibration record
Annexure-IIAuto titrator calibration record
Annexure-IIIElectrodes calibration record-pH
Annexure-IVElectrodes calibration record-Solvotrode
Annexure-VElectrodes calibration record-Combined Ag ring

         ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
SOP:Standard Operating Procedure
GTP:General Test Procedure
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Annexure-I
BURETTE CALIBRATION RECORD

BURETTE CALIBRATION RECORD

Annexure-II
AUTO TITRATOR CALIBRATION RECORD

Annexure-III
ELECTRODE CALIBRATION RECORD-pH Electrode

ELECTRODE CALIBRATION RECORD-pH Electrode

Annexure-IV
ELECTRODE CALIBRATION RECORD-SOLVOTRODE

ELECTRODE CALIBRATION RECORD-SOLVOTRODE

Annexure-V
ELECTRODE CALIBRATION RECORD-Combined Ag ring

ELECTRODE CALIBRATION RECORD-Combined Ag ring
Share this article
Shareable URL
Prev Post

OPERATION OF VACUUM PUMP

Next Post

CONDITIONS FOR STOPPING STABILITY STUDY

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading