CALIBRATION OF B.O.D INCUBATOR CHAMBERS

  • PROCEDURE:
  • Keep all sensors of calibrator for temperature monitoring inside the BOD incubator chamber near to the existing inbuilt sensors of chamber as per the Format-I.
  • Switch on the calibrator and observe the display for temperature.
  • Calibrator should be standardized at ‘NIST’ or ‘ERTL’ or ‘NPL’ or ‘ETDC’ equivalent authorities and the traceability certificates shall be obtained.
  • The calibration may be done by the external agency if required, as per the procedure mentioned in this SOP.
  • Calibration duration shall be minimum of 4 hrs. for the set condition.
  • Readings shall be taken with 15 minutes time intervals.
  • Calibration shall be carried out on ±15 days, from the calibration due date.
  • After the calibration record the results in calibration record as per the Format-II.
  • Calibration Frequency: Once in year.
  • Acceptance criteria: ± 1 0C for set temperature.

Note: Attach the data logger printouts with Format-I.

Click the link to download word file copy of this document: https://pharmaguidehub.com/product/calibration-of-b-o-d-incubator-chambers/

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IIncubator layout for probe placement.
Annexure-IIIncubator Chambers Calibration Record
  • ABBREVIATIONS:
No.:Number
SOP:Standard Operating Procedure
ERTL:Electronic Regional Testing Laboratory
NIST:National institute of standards and testing
NPL:National Physical Laboratory.
ETDC:Electronic Test & development Centre
QC:Quality Control

Annexure-I
Incubator layout for probe placement

Incubator layout for probe placement

Annexure-II
Incubator Chambers Calibration Record

Incubator Chambers Calibration Record
Incubator Chambers Calibration Record

Click the link to download word file copy of this document: https://pharmaguidehub.com/product/calibration-of-b-o-d-incubator-chambers/

Share this article
Shareable URL
Prev Post

CALIBRATION OF ACCURACY, PRECISION, LINEARITY AND SYSTEM SUITABILITY OF TOC ANALYZER

Next Post

CALIBRATION OF COLONY COUNTER (MAKE: CINTEX)

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading