- PROCEDURE:
- LABORATORY INCUBATOR
- Make: NEWTRONIC
- Keep the calibrated thermometer in the center of the Incubator.
- Set the temperature controller knob at 57.5 ℃, wait until the temperature is stable and Measure the temperature with calibrated thermometer.
- Note down the readings in calibration record as per Format -I.
- The tolerance limit is ± 2°C from the set temperature.
- If the calibration is not proper report the results to the departmental head for taking an appropriate action.
- Frequency of Calibration is Half yearly
- ANNEXURES:
ANNEXURE NO. | TITLE OF ANNEXURE |
Annexure-I | Calibration Record of Laboratory Incubator |
Annexure-I
CALIBRATION RECORD OF LABORATORY INCUBATOR