CALIBRATION OF LABORATORY INCUBATOR (Make: NEWTRONIC)

  • PROCEDURE:
  • LABORATORY INCUBATOR
  • Make: NEWTRONIC
  • Keep the calibrated thermometer in the center of the Incubator.
  • Set the temperature controller knob at 57.5 ℃, wait until the temperature is stable and Measure the temperature with calibrated thermometer.
  • Note down the readings in calibration record as per Format -I.
  • The tolerance limit is ± 2°C from the set temperature.
  • If the calibration is not proper report the results to the departmental head for taking an appropriate action.
  • Frequency of Calibration is Half yearly
  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-ICalibration Record of Laboratory Incubator

Annexure-I
CALIBRATION RECORD OF LABORATORY INCUBATOR

CALIBRATION RECORD OF LABORATORY INCUBATOR