- PROCEDURE:
- LABORATORY OVEN
- Make: HBio
- Model: BGP-9140A
- Operate the instrument as per operation SOP.
- Set the temperature at 50°C, wait until the temperature is stable and measure the temperature with calibrated thermometer.
- Note down the readings in calibration record as per Format-I.
- Repeat the same at temperature 100°C.
- The tolerance limit is ± 2°C from the set temperature.
- If the calibration is not proper report the results to the departmental head for taking an appropriate action.
- Frequency of Calibration is Half yearly.
- ANNEXURES:
ANNEXURE NO. | TITLE OF ANNEXURE |
Annexure-I | Calibration record of Laboratory Oven |
Annexure-I
Calibration Record of Laboratory Oven