CALIBRATION OF LABORATORY OVEN Make: HBio

  • PROCEDURE:
  • LABORATORY OVEN
  • Make: HBio
  • Model: BGP-9140A
  • Operate the instrument as per operation SOP.
  • Set the temperature at 50°C, wait until the temperature is stable and measure the temperature with calibrated thermometer.
  • Note down the readings in calibration record as per Format-I.
  • Repeat the same at temperature 100°C.
  • The tolerance limit is ± 2°C from the set temperature.
  • If the calibration is not proper report the results to the departmental head for taking an appropriate action.
  • Frequency of Calibration is Half yearly.
  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-ICalibration record of Laboratory Oven

Annexure-I
Calibration Record of Laboratory Oven

Calibration record of Laboratory Oven