- OBJECTIVE:
To lay down the procedure for Calibration of muffle furnace.
- SCOPE:
This SOP is applicable to the procedure for Calibration of muffle furnace at {Company Name} {Location}.
- RESPONSIBILITY:
- Executive/Designee Quality Control is responsible to calibrate the instrument as per SOP.
- Head QC: Ensure the compliance of SOP.
- ACCOUNTABILITY:
QA Head shall be Accountable for implementation of SOP.
- PROCEDURE:
- Muffle furnaces are incredibly versatile pieces of equipment used to generate high temperatures in a controlled environment.
- Operate the instrument as per SOP.
- Set the temperature to 600°C, 800°C & 1000°C by pressing `▲` or `▼`, wait until the temperature is stable and measure the temperature with calibrated thermocouple.
- Note down the observed readings in calibration record as per Format-I.
- The tolerance limit is ± 25°C from the set temperature.
- If the calibration is not proper report the results to the department head for taking an appropriate action.
- Frequency of Calibration: Half yearly.
- REFERENCES:
Not Applicable
- ANNEXURES:
ANNEXURE NO. | TITLE OF ANNEXURE |
Annexure-I | Calibration record of Muffle furnace |
ENCLOSURES: SOP Training Record.
- DISTRIBUTION:
- Controlled Copy No. 01 : Head Quality Assurance
- Controlled Copy No. 02 : Head Quality Control
- Master Copy : Quality Assurance Department
- ABBREVIATIONS:
No. | : | Number |
SOP | : | Standard Operating Procedure |
QC | : | Quality Control |
- REVISION HISTORY:
CHANGE HISTORY LOG
Revision No. | Details of Changes | Reason for Change | Effective Date |
00 | New SOP | Not Applicable | To Be Written Manual |
Annexure-I
CALIBRATION RECORD OF MUFFLE FURNACE