Your Compass for Pharma Success

CALIBRATION OF TAP DENSITY APPARATUS

  • OBJECTIVE:
  • To lay down the procedure for Calibration of Tap density apparatus.
  • SCOPE:
  • This SOP is applicable to the procedure for Calibration of Tap density apparatus at {Company Name} {Location}.
  • RESPONSIBILITY:
  • Executive/Designee Quality Control is responsible to calibrate the instrument as per SOP.
  • Head QC: Ensure the compliance of SOP.
  • ACCOUNTABILITY:

             QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:
    • Before switching on the instrument, check the area and instrument cleanliness.
    • Connect the plug to the power supply and switch on the power.
    • Operate the instrument as per SOP.
    • The following parameters to be calibrated for Tap density apparatus:
  • Drop height of cylinder holder by using digital vernier caliper
  • Number of drops per minute by using stopwatch
    • Calibration of Drop height of cylinder holder for USP method 1:
      • Open the top cover by removing bottom screws.
      • Turn the cam attached to the motor shaft clock wise, so that the cylinder holder shaft is at the minimum position.
      • Make sure that the bottom plate of the cylinder holder is in perfect contact with the tapping platform.
      • Measure the distance between the top plate of frame and the clamp plate of cylinder holder by using digital vernier caliper; this reading is the base height (HR) as shown in figure (a).
      • Rotate the cam further by hand, clockwise till the cylinder holder shaft reaches maximum height.
      • At this position the shaft tip should be on the falling edge of the cam arm. Mark this arm No; 1 as shown in figure (b).
      • Measure the distance between the top plate of frame and clamp plate of cylinder holder by using digital vernier caliper. This is height HI.
  • Make a mark on the point of measurement on the top plate of frame and on the cylinder plate.
    • The difference between these two readings is drop height of the cylinder.
    • Mark arms 1, 2 and 3 and heights H1, H2, H3 and so on. Repeat the above procedure for all arms with both cylinder holders.
    • Make sure that the point of measurement on the top plate of frame and on the cylinder holder plate is maintained the same for all measurements.
    • Record the results in the calibration format as per Format-I.
  • Repeat the same procedure for all arms with both cylinder holder (figure c) and record the results in the calibration format.
    • Calibration for Number of drops of cylinder holder per minute:
      • Choose the mode menu to ‘USP1’ by using digital scroll key.
      • Press set key, set taps count to 300.
      • Press start key to start the tapping and start the stop watch.
      • Stop the stop watch after 1 Minute and note down the number of drops.
      • The number of taps should be 300 ± 6.
      • Similarly choose the mode menu to ‘USP11’ by using digital scroll key.
      • Press set key, set taps count to 250.
      • Press start key to start the tapping and start the stop watch.
      • Stop the stop watch after 1 Minute and note down the number of drops.
      • The number of drops should be 250 ± 5.
  • REFERENCES:

Not Applicable

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-ITap density apparatus Calibration Record

        ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
SOP:Standard Operating Procedure
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

               Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Annexure-I
TAP DENSITY APPARATUS CALIBRATION RECORD

Share this article
Shareable URL
Prev Post

CALIBRATION OF REFRACTOMETER (MAKE – RUDOLPH RESEARCH)

Next Post

TESTING AND RELEASE OF RAW MATERIALS

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading