Your Compass for Pharma Success

CALIBRATION OF VACUUM OVEN MAKE: NEWTRONIC

  • OBJECTIVE:
  • To lay down the procedure for Calibration of Vacuum Oven, Make: Newtronic.
  • SCOPE:

This SOP is applicable to the procedure for Calibration of Vacuum Oven at {Company Name} {Location}.

  • RESPONSIBILITY:
  • Executive/Designee Quality Control is responsible to calibrate the instrument as per SOP.
  • Head QC: Ensure the compliance of SOP.
  • ACCOUNTABILITY:

             QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:

A vacuum oven is a specialized oven that removes moisture and other gases from a substance by creating a low-pressure environment inside the chamber. This is different from a standard oven, which uses heat and air circulation to dry materials.

  • Operate the instrument as per SOP.
    • Set the temperature controller knob to 60°C.
    • After the set temperature is attained, measure it with a calibrated thermometer.
    • The observed temperature should be within ± 2.0°C.
    • Similarly perform the test at 100°C and 150°C and measure the temperature with a calibrated thermometer.
    • The observed temperature should be within ± 2.0°C.
    • Record the readings in vacuum oven calibration record as per Format-I.
    • Create a vacuum of 400 mm and close the valve and observe the Vacuum drop for 3 hours. The Vacuum drop should be not more than 1.0% of the initial Pressure after 3 hours.
    • The frequency for calibration of vacuum oven is Six months.
    • If the calibration is not proper, report the results to the departmental head for taking an appropriate action.
  • REFERENCES:

Not Applicable

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IVacuum Oven Calibration record

        ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
SOP:Standard Operating Procedure
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Annexure-I
VACUUM OVEN CALIBRATION RECORD

Share this article
Shareable URL
Prev Post

CALIBRATION OF MELTING POINT APPARATUS

Next Post

CALIBRATION OF ALLIANCE HPLC WITH FLUORESCENCE DETECTOR

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading