OPERATION OF ELECTRICALLY OPERATED AUTOMATIC PRESS

  • PROCEDURE:
  • Make: Techno search Instruments, Model: AP-15
  • Ensure that the instrument is clean and free from dust, Traces of Oil etc.
  • Ensure that the power cord is connected to the power socket.
  • ‘ON’ the main switch fitted on the front side of the press.
  • Clean the Anvil and plunger with solvent to remove the grease.
  • Insert the Anvil with polished surface facing with upwards in the SS die.
  • Fill evenly the powder sample on the Anvil surface and spread it using plunger.Insert plunger facing polished surface downwards.
  • Place the assembled Die on the cover plate of cylinder of the press.
  • Move the screw with handle turning clockwise until it touches the plunger.
  • Close the release knob turning clockwise.Insert acrylic sheets on front and back side slots.
  • Apply pressure by pressing ‘START’ switch as the pumping operation starts by electrically(normally pressure required 8-10 tons).
  • The pressure will stop as soon as the pressure reaches to set value in the pressure switch.
  • Wait for appropriate time.Release the pressure by turning the release knob in anticlockwise direction.
  • Take out the total die from press and keep again the same assembled die in opposite direction to remove the pellet (anvil upward and plunger downward).
  • Keep the SS stand on anvil, touch the screw by handle in clockwise direction (manually).
  • Clean all the traces of sample and store the Die in Dry Box or Desiccator.
Share this article
Shareable URL
Prev Post

OPERATING PROCEDURE OF HOT WATER GENERATOR

Next Post

CALIBRATION OF ALLIANCE HPLC WITH EMPOWER SOFTWARE

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading