OPERATION OF PIPETTES WASHER

  • OBJECTIVE:
  • To lay down the procedure for operation of the pipettes washer in QC Department.
  • SCOPE:
  • This SOP is applicable to the procedure for Operation of the pipettes washer in QC Department at {Company Name} {Location}.
  • RESPONSIBILITY:
  • Officer/Executive/Designee Quality Control – Shall be responsible for operation as per SOP.
  • Head/Designee Quality Control – Shall be responsible for ensuring compliance as per SOP.
  • ACCOUNTABILITY:
  • QA Head shall be Accountable for implementation of SOP.
  • PROCEDURE:
  • Keep all the washable pipettes into the pipettes washer Apparatus.
  • Fill the Apparatus above the level of the pipettes with raw water and move the sliding mechanism Up and down for five times then drain the water.
  • Add approximately (25ml) of Teepol/Equivalent solution into the washer and fill the washer with Raw water above the level of the pipettes.
  • Move the sliding mechanism up and down of the Pipettes washer for 5 times then drain the water.
  • Fill the apparatus with Purified water above the level of the pipettes and move the sliding Mechanism up and down for five times.
  • Wash the pipettes till the Soap solution is removed completely.
  • Take out all the pipettes one by one and clean with Acetone, keep for drying at 60 Deg C.
  • Take out the dried pipettes from the oven and keep in their designated places.

Click the link to download word file copy of this document:
https://pharmaguidehub.com/product/operation-of-pipettes-washer/

REFERENCES:

Not Applicable

ANNEXURES: 

Not Applicable

        ENCLOSURES: SOP Training Record.

 DISTRIBUTION:

  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department

ABBREVIATIONS:

No.:Number
SOP:Standard Operating Procedure
QC:Quality Control

REVISION HISTORY:            

CHANGE HISTORY LOG
Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Click the link to download word file copy of this document:
https://pharmaguidehub.com/product/operation-of-pipettes-washer/

Share this article
Shareable URL
Prev Post

OPERATION OF pH METER

Next Post

OPERATION OF VACUUM OVEN

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading