Your Compass for Pharma Success

OPERATION OF UV VISIBLE SPECTROPHOTOMETER

  • OPERATING PROCEDURE OF UV VISIBLE SPECTROPHOTOMETER :
  • Make: SHIMADZU UV-180
  • Start-up procedure:
  • Turn on instrument.
  • Allow instrument to warm up for fifteen minutes.
  • Turn on the computer and login.
  • Configure a network printer and set it as default.
  • Make sure you are connected to the network.
  • Open UV Probe 2.34 (Start>Programs>Shimadzu>UV Probe or Select Desktop Icon)
  • Control of instrument to computer:On the instrumentAfter initialization of instrument, the mode screen will be displayed.
  • Press F4 key (PC Control) on the instrument keypad to control through the computer.
  • On the computer Connect to instrument (Instrument >Commends >Connect)
  • Create /Edit Method:
  • Load method (Edit>Method)
  • To create new method select spectrum module icon in menu bar.
  • In Edit Method parameters,
  • Select Measurement Tab.
  • Set the Wavelength range –starting and ending values
  • Set scan speed.
  • Set auto sampling interval.
  • Under Scan Mode click auto on the file name icon and give product Analytical Reference Number as file name.
  • Instrument parameters Tab.
  • Select absorbance as measurement mode.
  • Click Ok
  • Change Y-axis range.
  • Go to the overlay (select tab at top of graph or Graph>overlay) and click on the minimum and maximum absorbance values.
  • Enter a new value
  • When finished, click anywhere on the screen to get out of this mode
  • Save method (File > Save as)
  • Enter file name.
  • In the Save as Type list, click Method file (*.smd)
  • Click Save
  • Collect spectrum
  • Perform baseline correction
  • Make sure sample and reference compartments are clear.
  • Place the two UV cells with blank solution.
  • Collect baseline (Instrument>Commands>Baseline).  
  • When the Baseline parameter box appears, enter the starting and ending values.
  • Click OK.
  • Insert sample (front cell holder) and reference (back cell holder).
  • Start (Instrument >Commands> Start).
  • When scan is complete, save (File >save All).
  • Remove sample and reference from compartment.
  • Analyze spectrum:
  • Perform peak pick (Operation >peak pick)
  • The operation screen will open to display results.
  •  Right click on the peak pick table and make sure mark peaks. and show peaks are selected.
  • Adjust peak threshold (if necessary )
  • Right click on peak pick table and select properties.
  • Enter new threshold value.
  • Close Operation screen (view operation).
  • Print peak pick results (File>print>Current View).
  • Print Spectrum:
  • Select report icon from menu bar and select spectrum report format.Print spectrum (File >print).
  • For Photometric measurement
  • Load method (Edit >Method)
  • To create new method, select photometric module icon in menu bar.
  • In Edit method parameter parameters,
  • Select measurement Tab.
  • Set wavelength type and wavelength (nm)
  • Click on add Button for wave length and click on next button.
  • Select type formula, units, Wavelength and Standard concentration and Click on next button for properties.
  • Click on file name side icon and give product AR No as a file name. title is product name comments are Batch Number
  • Click on finish button
  • Save method (file >save as).
  • Enter file name.
  • In the save type list, click method File (*.smd).
  • Click on the save icon for save file.
  • Collect photometric data:
  • Perform base line correction.
  • Make sure sample and reference compartments are clear.
  • Place two UV cells with a blank solution.
  • Click AUTO ZETO Replace the blank solution in sample cell with sample solution Give sample ID in sample table (active) Click on the READ UNK button.
  • Wait for 20 seconds till the show a constant value After completion of measurement click on the save button the data will be saved Select the icon form menu bar and select the photometric report format
  • Take printout.
  • Enter the details in logbook.
  • Shutdown procedure:
  • Press disconnect on computer to disconnect from instrument (Instrument >commands >disconnect).
  • Turn off instrument.
  • Log off computer.
  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IInstrument usage logbook

Annexure-I

UV-VIS SPECTROPHOTOMETER
Share this article
Shareable URL
Prev Post

CALIBRATION OF WEIGHING BALANCES

Next Post

OPERATION OF PARTICLE SIZE ANALYZER

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading