PROCEDURE FOR MONITORING OF COMPRESSED AIR

  • PROCEDURE:
  • Connect the unit to the desired point of sample using sterile tubing. The inlet gas valve should be in a closed position.
  • Insert an identified petri dish with sterile medium SCD in the “Pinocchio funnel” housing and screw it to the conical head.
  • Open the line of Compressed air or gas line.
  • Open the gas valve ‘D’ + ‘D1’ and adjust the airflow on flow meter to 100 Liters / min (210 SCFH) to wash the line.
  • Connect the “Pinocchio funnel” to the “Pinocchio Super”.
  • Close the ‘D1’ valve and open ‘D2’ to start the countdown.
  • Close the gas valve after 10 minutes (1000 liters of air is collected).
  • Unscrew the head and transfer the petri dish to the laboratory for incubation.
  • The “Pinocchio Super” is ready for a new test.
  • Incubate the plates in the inverted position at 30-35°C for NLT 72 hours.
  • Count the colony forming units in the plate at the end of incubation time and report the result as cfu/1000 liters of gas/ compressed air.
  • After incubation count the colonies. If any colonies are observed streak on selective media.
  • Positive control and negative control for the respective media shall also be maintained.

LIMITS: 5 cfu/1000L; Pathogens: Enterobacteriaceae & Pseudomonas should be absent.

Frequency: Half yearly.

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IReport for Bio Load of Compressed air.
Annexure-IIPinocchio Super II-Scheme
  • ABBREVIATIONS:
No.:Number
cfu:Colony Forming Units
NLT:Not Less Than
SOP:Standard Operating Procedure
QC:Quality Control

Annexure-I
REPORT FOR BIOLOAD OF COMPRESSED AIR

REPORT FOR BIOLOAD OF COMPRESSED AIR

Annexure-II
Pinocchio Super II-Scheme

Pinocchio Super II-Scheme
Share this article
Shareable URL
Prev Post

MAINTENANCE OF ONLINE TOC ANALIZER (Model: 550, Make: METTLER TOLEDO)

Next Post

OPERATION OF LINOMAT 5 TLC SPOTTER MAKE: CAMAG

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

OPERATION OF UV CABINET

QA Head shall be Accountable for implementation of SOP. Click the link to download word file copy of this…

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading