WASHING OF HPLC COLUMNS

  • OBJECTIVE:
  • To lay down the procedure for washing of HPLC columns.
  • SCOPE:
  • This SOP is applicable to the procedure for the washing of HPLC columns at {Company Name} {Location}.
  • RESPONSIBILITY:
  • Quality Control Officer/Executive: is responsible to perform the activity as per the SOP
  • Head Quality Control: is responsible to ensure the compliance of this SOP
  • ACCOUNTABILITY:

             QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:
  • For C18, C8, C6, Phenyl, CN, amine columns:
  • Change the mobile phase to Milli Q water.Allow the water to flow through the column with flow rate of 0.5 ml increment at the rate of 1ml / min. for minimum 30 minutes.
  • Change the mobile phase from water to HPLC Methanol or Acetonitrile.
  • Set the instrument to flow rate 1 ml/min. and wash the column for 30 minutes.
  • In case of Reverse phase columns, ensure that the column contains methanol.
  • In case of normal phase columns, ensure that columns shall be washed and stored in Hexane.
  • Disconnect the column from the HPLC system and store in an appropriate place
  • For protein pack column:
  • Change the mobile phase to 10% Isopropyl Alcohol.
  • Allow 10% Isopropyl Alcohol to flow through the column with flow rate of 0.5 ml increment at the rate of 1ml/min. for minimum 30 minutes.
  • Ensure that the column contains 10% Isopropyl Alcohol.
  • Disconnect the column from the HPLC unit and store in an appropriate place.
  • For L17 column:
  • Change the mobile phase to a mixture of 0.01N H2SO4: ACN (85: 15).
  • Allow the mixture to flow through the column at the rate of 0.6ml/min for 2 hours.
  • Flow rate should not exceed 0.6ml/min.
  • Change the mobile phase from (H2SO4, ACN) to water.
  • Set the instrument to flow rate 0.5ml/min. and wash the column for 2 hours.
  • Ensure that the column contains water.
  • Disconnect the column from the HPLC unit and store it in an appropriate place.

Click the link for download word file copy of this document:
https://pharmaguidehub.com/product/washing-of-hplc-columns/

  • REFERENCES:

Not Applicable

  • ANNEXURES: 

Not Applicable

        ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
OOS:Out of specification
SOP:Standard Operating Procedure
HPLC:High Performance Liquid Chromatography
H2SO4:Sulphuric acid
ACN:Acetonitrile
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Click the link for download word file copy of this document:
https://pharmaguidehub.com/product/washing-of-hplc-columns/

Share this article
Shareable URL
Prev Post

TESTING AND RELEASE OF INPROCESS AND FINISHED PRODUCTS

Next Post

CLEANING OF DISINTEGRATION TEST APPARATUS (MAKE – ELECTRO LAB)

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading