- OBJECTIVE:
- To lay down the procedure for washing of HPLC columns.
- SCOPE:
- This SOP is applicable to the procedure for the washing of HPLC columns at {Company Name} {Location}.
- RESPONSIBILITY:
- Quality Control Officer/Executive: is responsible to perform the activity as per the SOP
- Head Quality Control: is responsible to ensure the compliance of this SOP
- ACCOUNTABILITY:
QA Head shall be Accountable for implementation of SOP.
- PROCEDURE:
- For C18, C8, C6, Phenyl, CN, amine columns:
- Change the mobile phase to Milli Q water.
- Allow the water to flow through the column with flow rate of 0.5 ml increment at the rate of 1ml / min. for minimum 30 minutes.
- Change the mobile phase from water to HPLC Methanol or Acetonitrile.
- Set the instrument to flow rate 1 ml/min. and wash the column for 30 minutes.
- In case of Reverse phase columns, ensure that the column contains methanol.
- In case of normal phase columns, ensure that columns shall be washed and stored in Hexane.
- Disconnect the column from the HPLC system and store in an appropriate place
- For protein pack column:
- Change the mobile phase to 10% Isopropyl Alcohol.
- Allow 10% Isopropyl Alcohol to flow through the column with flow rate of 0.5 ml increment at the rate of 1ml/min. for minimum 30 minutes.
- Ensure that the column contains 10% Isopropyl Alcohol.
- Disconnect the column from the HPLC unit and store in an appropriate place.
- For L17 column:
- Change the mobile phase to a mixture of 0.01N H2SO4: ACN (85: 15).
- Allow the mixture to flow through the column at the rate of 0.6ml/min for 2 hours. Flow rate should not exceed 0.6ml/min.
- Change the mobile phase from (H2SO4, ACN) to water.
- Set the instrument to flow rate 0.5ml/min. and wash the column for 2 hours.
- Ensure that the column contains water.
- Disconnect the column from the HPLC unit and store it in an appropriate place.
- For C18, C8, C6, Phenyl, CN, amine columns:
- REFERENCES:
Not Applicable
- ANNEXURES:
Not Applicable
ENCLOSURES: SOP Training Record.
- DISTRIBUTION:
- Controlled Copy No. 01 : Head Quality Assurance
- Controlled Copy No. 02 : Head Quality Control
- Master Copy : Quality Assurance Department
- ABBREVIATIONS:
No. | : | Number |
OOS | : | Out of specification |
SOP | : | Standard Operating Procedure |
HPLC | : | High Performance Liquid Chromatography |
H2SO4 | : | Sulphuric acid |
ACN | : | Acetonitrile |
QC | : | Quality Control |