WASHING OF HPLC COLUMNS

  • OBJECTIVE:
  • To lay down the procedure for washing of HPLC columns.
  • SCOPE:
  • This SOP is applicable to the procedure for the washing of HPLC columns at {Company Name} {Location}.
  • RESPONSIBILITY:
  • Quality Control Officer/Executive: is responsible to perform the activity as per the SOP
  • Head Quality Control: is responsible to ensure the compliance of this SOP
  • ACCOUNTABILITY:

             QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:
    • For C18, C8, C6, Phenyl, CN, amine columns:
      • Change the mobile phase to Milli Q water.
      • Allow the water to flow through the column with flow rate of 0.5 ml increment at the rate of 1ml / min. for minimum 30 minutes.
      • Change the mobile phase from water to HPLC Methanol or Acetonitrile.
      • Set the instrument to flow rate 1 ml/min. and wash the column for 30 minutes.
      • In case of Reverse phase columns, ensure that the column contains methanol.
      • In case of normal phase columns, ensure that columns shall be washed and stored in Hexane.
      • Disconnect the column from the HPLC system and store in an appropriate place
    • For protein pack column:
      • Change the mobile phase to 10% Isopropyl Alcohol.
      • Allow 10% Isopropyl Alcohol to flow through the column with flow rate of 0.5 ml increment at the rate of 1ml/min. for minimum 30 minutes.
      • Ensure that the column contains 10% Isopropyl Alcohol.
      • Disconnect the column from the HPLC unit and store in an appropriate place.
    • For L17 column:
      • Change the mobile phase to a mixture of 0.01N H2SO4: ACN (85: 15).
      • Allow the mixture to flow through the column at the rate of 0.6ml/min for 2 hours. Flow rate should not exceed 0.6ml/min.
      • Change the mobile phase from (H2SO4, ACN) to water.
      • Set the instrument to flow rate 0.5ml/min. and wash the column for 2 hours.
      • Ensure that the column contains water.
      • Disconnect the column from the HPLC unit and store it in an appropriate place.
  • REFERENCES:

Not Applicable

  • ANNEXURES: 

Not Applicable

        ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
OOS:Out of specification
SOP:Standard Operating Procedure
HPLC:High Performance Liquid Chromatography
H2SO4:Sulphuric acid
ACN:Acetonitrile
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual
Share this article
Shareable URL
Prev Post

TESTING AND RELEASE OF INPROCESS AND FINISHED PRODUCTS

Next Post

CLEANING OF DISINTEGRATION TEST APPARATUS (MAKE – ELECTRO LAB)

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading