OPERATION OF REFRACTOMETER (MAKE – RUDOLPH RESEARCH)

  • OBJECTIVE:
  • To lay down a procedure for operation of Refractometer, Make – Rudolph Research.
  • SCOPE:

This SOP is applicable to the procedure for operation of Refractometer, Make – Rudolph Research at {Company Name} {Location}.

  • RESPONSIBILITY:
  • Officer/Executive/Designee Quality Control – Shall be responsible for operation as per SOP.
  • Head/Designee Quality Control – Shall be responsible for ensuring compliance as per SOP.
  • ACCOUNTABILITY:

             QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:
    • Check the cleanliness of the work area and instrument.
    • Connect the power cord to the electrical supply source.
    • Switch on the instrument by using the ON/OFF switch present on the backside of the instrument.
    • Lift the presser and clean the sample well by using solvents like water, ethanol, acetone, Iso propyl alcohol.
    • Place the sample to be measured in the sample well, after placing the sample in the sample well, close the presser.
    • Place enough sample so that it touches the inner part of the silver metal ring all the way round.
    • Now press the Temp soft key present on the touch screen, two options are displayed
  • Temp Control
  • Temp Correct
    • Now press the Temp control soft key and enter the required temperature.
    • Press the Measure key, system prompts for operator password enter the password and press enter.
    • Now press continue the stabilizing temperature message displays.
    • Press Enter manual and enter the Lot ID and press enter.
    • Press Enter manual and enter the sample ID and press enter.
    • The sample Refractive index measurement takes place.
    • Once the scan is complete the results are displayed.
    • Now press Yes key if results are accepted, No key if results are not accepted.
    • Now the system prompts for Review electronic signature enter the password, press continue and enter the comment and press enter.
    • Accept this comment message displays, press Yes if comment is accepted, No if comment is not accepted.
    • If No is pressed then enter the new comment and press enter, accept this comment message displays, press enter.
    • 21CFR11 back up in process message is displayed and the data backup is completed automatically.
    • Now press the Print/Send key and the report is printed.
    • Once the analysis is completed wipe off the sample with paper towel, clean the sample well using water or suitable solvent depending upon the nature of sample.
    • Enter the details of usage in the logbook.
  • REFERENCES:

Not Applicable

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IRefractometer User Logbook

         ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
SOP:Standard Operating Procedure
NMT:Not More Than
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Annexure-I
Refractometer User Logbook

Share this article
Shareable URL
Prev Post

QUALIFICATION OF ANALYST IN QUALITY CONTROL

Next Post

OPERATION OF DISINTEGRATION TEST APPARATUS (MAKE – ELECTRO LAB)

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

OPERATION OF pH METER

ANNEXURE NO. TITLE OF ANNEXURE Annexure-I pH Meter user logbook. Annexure-I pH METER USER LOGBOOK

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading