Your Compass for Pharma Success

OPERATION OF TAP DENSITY APPARATUS

  • PROCEDURE:
  • Before starting the operation check for the area cleanliness and instrument cleanliness.
  • Connect the pin to the power supply.
  • Switch on the power.
  • The instrument displays the Main Menu.
  • Choose the mode menu either ‘USP or User’ by pressing Digit scroll key.
  • Press the method key to set USP 1 or USP II.
  • CONDITION I (If as per USP 1)
  • Stroke Height: 14mm ± 2 mmStrokes/Min: Nominal 300 times
  • Tap count 1 : 500 times
  • Tap count 2 : 750 times
  • Tap count 3 : 1250 times
  • Press set key, to set tap count 1 to 500 by using Digital scroll and increase/decrease key.
  • Press enter key, set tap count 2 to 750 by using Digital scroll and increase/decrease key.
  • Press enter key, set tap count 3 to 1250 by using Digital scroll and increase / decrease key.
  • Press enter key, set clock time using Digital scroll and increase / decrease key.
  • Press enter key, set date by using shift key and increase / decrease key.
  • Press enter key, it will ask for “Print last result “if you need shift cursor to Y otherwise to N.
  • Press enter key.Press reset key, now your method get saved as per USP 1.
  • Press start key, input weight of sample by using shift key and increase / decrease key in grams.
  • Press enter key, input initial volume (Vo) by using shift key and increase / decrease key in ml.
  • Keep the cylinder in Holder and press start key; it will give a taps for 500 times.
  • Measured tapped volume (Va) and input by using shift key and increase / decrease key.
  • Press enter key and press start key. It will give a taps for 750 times.
  • Measured tapped volume (Vb) input by using shift key and increase / decrease key.
  • Note: If the difference volume of 1 st tap and 2 nd tap i.e. Va – Vb is more than 2% then proceed for tap count 3 i.e. 1250 times by same manner. Measure the final volume (Vf)
  • If the difference volume is not more than 2% stop the experiment in Tap count 2 only.
  • Give the connection to printer and go to printing mode by using SET and Enter Key.
  • Set printing mode to Y, automatically printing will come.
  • We can get tap density Initial Density, compressibility Index and Hausner Ratio.
  • CONDITION II (if as per USP II):
  • Stroke height: 3mm+/-10 %
  • Strokes/min: Nominal 250 times
  • Tap count 1: 500 times
  • Tap count 2: 750 times
  • Tap count 3: 1250 times
  • Proceed same as for condition 1.
  • CONDITION III (if as per user mode):
  • Set the mode to ‘User’ by using digital scroll keys.
  • Press set key, set taps count as per the product specification.
  • Press start key to start the tapping.
  • Measured the volume and calculate manually.
  • Calculate the bulk density of the sample according to the formula given below.
Bulk Density Formula
  • Switch off the instrument after usage.
  • Enter the details of usage in the logbook.
  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-ITap density user log

Annexure-I

TAP DENSITY USER LOG
Share this article
Shareable URL
Prev Post

OPERATION OF PARTICLE SIZE ANALYZER

Next Post

OPERATION OF pH METER

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading