Your Compass for Pharma Success

CALIBRATION OF GAS CHROMATOGRAPHY WITH EMPOWER SOFTWARE

  • PROCEDURE:
  • CALIBRATION OF FLOW ACCURACY:
  • Select detector flow and set it to the specified value listed in the result Table (Annexure-I)
  • Ensure that all other detector flows are “OFF”.After completing appropriate preparation steps for the calibration mass flow meter being used ( i.e. Gas selection, purge, and zero).
  • Connect the flow meter to the detector exit and measure flow.
  • Allow time for flow meters” reading to stabilize.
  • Record the flow value displayed on GC front panel in the results table.
  • Record the measured flow value displayed on flow meter in results table (Annexure-I).
  • Calculate the flow difference between displayed value and measured value and record the results table (Annexure-I).
  • Turn flow OFF.Repeat the steps above for each remaining detector flow.After all flows have been measured.
  • Remove the flow Cap from the detectors base.
  • Acceptance criteria: the difference between measured flow and set flow should not be more than ± 10%.
  • Record the details in annexure-I.
  • Calibration schedule: Once in six months and after any major maintenance Job.
  • CALIBRATION OF HEADSPACE SAMPLER WITH FID DETECTOR:
  • Fix the column from capillary inlet to FID.
  • Chromatographic parameters
  • Column: Fused silica column 30 m long 0.32 m internal diameter with 1.8 micrometers film of 6% Cynopropyl Phenyl. 94% dimethyl polysiloxane.
  • Inlet temperature 120°C
  • Carrier gas: Helium
  • Oven temperature: 60°C (Iso thermal)
  • Flow mode: Constant Flow
  • Detector Temperature: 200°C
  • Column Flow: 0.8 ml/min
  • Hydrogen Flow: 45 ml/min
  • Air flow :450ml/min
  • Make up flow :30ml/min
  • Mode: Split
  • Split ratio: 1:50
  • Stop time: 15 min.

Head space parameters

  • Incubation temperature: 80°C
  • Agitator Speed: 500 rpm
  • Incubation Time:20 min
  • Agitator on time: 10 sec
  • Agitator off time: 10 sec
  • Flush Time: 2.0 min
  • Syringe temperature: 90 °C
  • Pressurization Time: 0.08 min
  • Injection Volume: 1000 µL
  • Fill speed: 500 µl /sec
  • Fill stroke: 1
  • Injection speed: 500 µL/sec
  • GC Cycle Time :20 min
  • Sample preparation: prepare 0.4 % v/v of Ethanol in water.
  • Prepare six vials each contains 2 ml test solution. Seal the vial with PTFE septa and cap, place in the headspace analyzer. Run the experiment.
  • Retention time for ethanol shall be about 4.0 min.
  • Acceptance criteria:
  • RSD for peak area should not be more than 5.0 %
  • RSD for retention time should not be more than 0.5%
  • Record the details in Annexure-II.
  • Calibration Schedule: Once in six months and after any major maintenance Job.
  • CALIBRATION OF CAPILLARY INLET WITH FID:
  • Fix the column for capillary inlet to FID.
  • Chromatographic parameters
  • Column: HP –5, 30 mx0.32 mm ID x 0.25µ
  • Inlet temperature: 250°C
  • Carrier gas: Helium/Nitrogen
  • Flow mode: Constant pressure
  • Inlet pressure column 25 psi
  • Detector Temperature: 300°C
  • Injection Volume: 1 µL
  • Column Flow: 0.8 ml/min
  • Hydrogen Flow: 45 ml/min
  • Mode: Split less
  • Air Flow: 450 ml/min
  • Makeup Flow: 7 ml/min
  • Run time: 8.5 min
  • Oven temperature Program: Initially held at 60°C then raised to 90°C at 25°C/min. Then again increased to 170°C at 15°C /min. Hold at 170° C for 2 min.
  • Procedure: Inject the sample 6 times as per the procedure and calculate the RSD for retention times and peak areas of individual peaks.
  • Acceptance criteria:
  • RSD for peak area should not be more than 5.0 %
  • RSD for retention time should not be more than 0.5%
  • The average area of individual peaks should be greater than the value mentioned in the annexure-III.
  • Record the details in calibration record in Annexure-3
  • Calibration Schedule: Once in six months and after any major maintenance Job.
  • CALIBRATION OF DETECTOR LINEARITY FOR FID:
  • Chromatographic parameters:
  • Column: Fused silica column 30 m long 0.32 m internal diameter with 1.8µ
  • Injection temperature 190°C
  • Carrier gas: Helium/Nitrogen
  • Flow mode: Constant Flow
  • Detector Temperature: 250°C
  • Hydrogen Flow: 45 ml/min
  • Air Flow: 450ml/min.
  • Make flow :30ml/min
  • Column Flow: 1mL/min
  • Mode: Split
  • Split ratio: 1:50
  • Injection Volume: 1000 µL
  • Run time 20 min.
  • Oven temperature Program: Initially held at 70 °C for 10 min. Then raised to 260°C at 70°C/min and Hold at 260° C for 5 min.

Head space parameters

  • Incubation temperature: 120°C
  • Agitator Speed : 500 rpm
  • Incubation Time:30 min
  • Agitator on time : 10 sec
  • Agitator off time: 10 sec
  • Flush Time : 2.0 min
  • Syringe temperature : 130 °C
  • Pressurization Time: 0.08 min
  • Injection Volume :1000 µL
  • Fill speed : 500 µl /sec . Fill stroke: 1
  • Injection speed: 500 µL/sec .
  • GC Cycle Time: 25 min

Preparation IPA stock solution :

  • Take 500µl of IPA into a 25ml flask having about 20 ml of NN-DMF and then dilute it to volume with NN-DMF and mix well.

Preparation of solutions for Linearity.

S. NoVolume of STD stock (ml)Dilute to 10 ml of NN- DMF
01110
02210
03310
04410
05510
06610
  • Pipette 5 ml of each solution into individual vials. Fit them with a septum and crimp cap and then seal it and inject.
  • Calculate the coefficient of correlation.
  • Acceptance criteria: Coefficient of correlation NLT 0.99.
  • Calibration: Once in six months and after any major maintenance Job.
  • Record the Details in annexure-IV.

CALIBRATION OF COLUMN OVEN TEMPERATURE TEST:

  • Set the oven temperature to 35°C and wait for at least 5 min to stabilize the oven, take the temperature reading from the data logger and record the temperature.
  • Set the oven temperature to 100 °C and wait for at least 5 min, after the oven gets Stabilized record the data logger reading in the annexure.
  • Set the oven temperature to 230 °C and wait for at least 5 min, after the oven gets Stabilized record the data logger reading in the annexure.
  • Acceptance criteria:
  • The difference at 35°C should not be more than 2°C from the set value.
  • The difference at 100°C should not be more than 4°C from the set value.
  • The difference at 230° C should not be more than 5°C from the set value.
  • Record the details in Annexure-V.
  • Calibration schedule: Once in six months and after any major maintenance Job
  • VERIFICATION OF DETECTOR AND INJECTOR TEMPERATURE:
    • Detector:
  • Set the detector temperature to 100°C and wait for at least 15 min and record the Temperature in Annexure -VI.
  • Repeat the above step for remaining temperatures 200°C and 300°C.
    • Injector:
  • Set the injector temperature to 100°C and wait for at least 15 min and record the temperature in Annexure-VI.
  • Repeat the above step for remaining temperatures 200°C and 300°C.
    • Acceptance criteria:
  • The difference at 100°C should not be more than 2°C from the set value.
  • The difference at 200°C should not be more than 2°C from the set value.
  • The difference at 300°C should not be more than 2°C from the set value.
    • Calibration schedule: Once in six months and after any major maintenance Job.
  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-ICalibration record of Flow accuracy
Annexure-IICalibration record for Headspace sampler With FID
Annexure-IIICalibration record for capillary inlet with FID
Annexure-IVCalibration record of Detector Linearity for FID
Annexure-VCalibration record of Column oven temperature Test
Annexure-VIVerification of Detector and Injector temperature

Annexure-I
CALIBRATION RECORD OF FLOW ACCURACY

CALIBRATION RECORD OF FLOW ACCURACY

Annexure-II
CALIBRATION RECORD FOR HEADSPACE SAMPLER WITH FID

CALIBRATION RECORD FOR HEADSPACE SAMPLER WITH FID

Annexure-III
CALIBRATION RECORD FOR CAPILLARY INLET WITH FID

CALIBRATION RECORD FOR CAPILLARY INLET WITH FID

Annexure-IV
CALIBRATION RECORD OF DETECTOR LINEARITY FOR FID

CALIBRATION RECORD OF DETECTOR LINEARITY FOR FID

Annexure-V
CALIBRATION RECORD OF COLUMN OVEN TEMPERATURE

CALIBRATION RECORD OF COLUMN OVEN TEMPERATURE

Annexure-VI
VERIFICATION OF DETECTOR AND INJECTOR TEMPERATURE

VERIFICATION OF DETECTOR AND INJECTOR TEMPERATURE
Share this article
Shareable URL
Prev Post

CALIBRATION OF ALLIANCE HPLC WITH EMPOWER SOFTWARE

Next Post

CALIBRATION OF ALLIANCE HPLC WITH EMPOWER SOFTWARE FOR PDA AND UV DETECTOR

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

HANDLING OF DESICCATOR

To lay down the procedure for handling and Maintenance of Desiccator. This SOP is applicable for the procedure…

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading