OPERATION OF COLONY COUNTER (MAKE: PROTOCOL)

Colony Counter in micro
OPERATION OF COLONY COUNTER (MAKE PROTOCOL)

OBJECTIVE:

To lay down the procedure for operation of colony counter (Make: Protocol).

SCOPE:

This SOP is applicable for the procedure for operation of colony counter (Make: Protocol) at {Company Name} {Location}.

RESPONSIBILITY:

  • In charge- Microbiology- is responsible to ensure compliance as per SOP.
  • Head/Designee Quality Control – Shall be responsible for ensuring compliance as per SOP.

ACCOUNTABILITY:

QA Head shall be Accountable for implementation of SOP.

What is a Colony Counter?

A colony counter is a laboratory instrument used to accurately count the number of bacterial or fungal colonies growing on an agar plate. It streamlines the counting process, reducing human error and saving time. Modern colony counters often use image analysis software to automatically detect and count colonies, further improving efficiency and accuracy. This tool is essential in microbiology research, food safety testing, and pharmaceutical development.

PROCEDURE:

Press the rear push button next to the power supply and switch on the monitor.

Now system is ready for counting the plate.

Place the plate under the camera.

The image of the plate will be grabbed automatically and displayed in the image pane.

If necessary, exclude any region within the frame where you do not want to count colonies, such as an area where colonies have coalesced and cannot be separated.

To perform a manual count, right-click in the image pane to display the image context menu and choose dialog box.

Count each colony by pressing with mouse on the displayed images on the screen.

After completing the count, the entry shall be made in Format-I.

Click the link for download word file copy of this document: https://pharmaguidehub.com/product/operation-of-colony-counter-make-protocol/

Plate validation:

To validate the colony counter by using the standard validation plate.

Remove all dust from the standard validation plate.

Place the validation plate on the protocol unit in the protocol system in place of the dish holder.

Make sure that the colonies are all within the frame boundary.

Count the colonies on the validation plate by pressing the spacebar.

If the count per frame is 100, the system is working correctly.

REFERENCES:

Not Applicable

ANNEXURES:

ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IUsage Log Sheet for Colony Counter

ENCLOSURES: SOP Training Record.

DISTRIBUTION:

Master Copy                    :          Quality Assurance Department

Controlled Copy No. 01  :          Head Quality Assurance

Controlled Copy No. 02  :         Head Quality Control (Micro.)

ABBREVIATIONS:

No.:Number
SOP:Standard Operating Procedure
QC:Quality Control

REVISION HISTORY:      

CHANGE HISTORY LOG
Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Annexure-I
USAGE LOG SHEET FOR COLONY COUNTER

USAGE LOG SHEET FOR COLONY COUNTER

Click the link for download word file copy of this document: https://pharmaguidehub.com/product/operation-of-colony-counter-make-protocol/

Share this article
Shareable URL
Prev Post

CALIBARTION OF MEDIA DISPENSER, MAKE: BIOTOOL

Next Post

CLEANING OF AIR SAMPLER (Make: MERCK)

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading