  • OBJECTIVE:
  • To lay down the procedure for operation of Milli-Q Water purification System.
  • SCOPE:

This SOP is applicable to the procedure for Operation of Milli-Q Water purification System at {Company Name} {Location}.

  • RESPONSIBILITY:
  • Officer/Executive/Designee Quality Control – Shall be responsible for operation as per SOP.
  • Head/Designee Quality Control – Shall be responsible for ensuring compliance as per SOP.
  • ACCOUNTABILITY:

             QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:
  • For General Cleaning:
  • Ensure that the power supply to the instrument is switched ‘OFF’ before cleaning.
  • Clean the instrument with a clean dry cloth every day.
  • Use clean cloth dipped in soap solution if necessary.
  • Care should be taken to remove the water with a dry cloth.
  • Operating Procedure:
  • Milli-Q water: 
  • This is ultrapure water that has been produced using a Millipore Milli-Q water purification system.
  • Milli-Q water purification systems: 
  • These are systems that are used to produce Milli-Q water. 
  • Switch on the main power switch provided on the backside of the instrument.
  • Move the “Point of use” trigger in a click way till the green bulb glowing starts to initiate the water purifying process.
  • The green bulb starts glowing.When the bulb is glowing, conductivity or resistivity of waters displays.
  • Then further move the “Point of use” trigger to start the dispensing of water.
  • After collecting the water click on the “Point of use” trigger to its original position.
  • Press the “OPERATE / STAND BY” button for at least 2 seconds to go to “STAND BY” mode.
  • The water resistivity shall be verified twice in a day.Instrument shall not be used if the output water resistivity is less than 18.2 MW. cm.
  • The Cartridges shall be replaced as below given and the details of the change shall be entered in Annexure-II.
  • Quantum Ultra-pure Cartridge: As and when required.Q-Guard Purification pack: As and when required.
  • Mill pack 40: When the flow rate is too low (<0.5 Liters per Minute)

Click the link for download word file copy of this document:
https://pharmaguidehub.com/product/operation-of-milli-q-water-purification-system/

  • REFERENCES:

Not Applicable

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IDaily Monitoring Record
Annexure-IICartridge Replacement record

         ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
SOP:Standard Operating Procedure
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

               Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Written Manual

Annexure-I
Daily Monitoring Record

Click the link for download word file copy of this document:
https://pharmaguidehub.com/product/operation-of-milli-q-water-purification-system/

Share this article
Shareable URL
Prev Post

OPERATION OF FRIABILATOR MAKE: ELECTROLAB

Next Post

ANALYTICAL WORK SHEET AND COA PREPARATION

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

OPERATION OF UV CABINET

QA Head shall be Accountable for implementation of SOP. Click the link to download word file copy of this…

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading