CALIBRATION OF LABORATORY OVEN

  • OBJECTIVE:
  • To lay down the procedure for Calibration of Laboratory oven.
  • SCOPE:

This SOP is applicable to the procedure for Calibration of Laboratory oven at {Company Name} {Location}.

  • RESPONSIBILITY:
  • Executive/Designee Quality Control is responsible to calibrate the instrument as per SOP.
  • Head QC: Ensure the compliance of SOP.
  • ACCOUNTABILITY:

             QA Head shall be Accountable for implementation of SOP.

  • PROCEDURE:
    • Operate the instrument as per SOP.
    • Set the temperature controller knob at 60 °C, wait until the temperature is stable and measure the temperature with calibrated thermometer.
    • Note down the readings in calibration record as per Format-I.
    • Repeat the same at temperatures 100°C and 150°C.
    • The tolerance limit is ± 2°C from the set temperature.
    • If the calibration is not proper report the results to the departmental head for taking appropriate action.
    • Frequency: of Calibration is Half yearly.

Click the link for download word file copy of this document:
https://pharmaguidehub.com/product/calibration-of-laboratory-oven/

  • REFERENCES:

Not Applicable

  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-ICalibration record of Lab Oven

        ENCLOSURES: SOP Training Record.

  •  DISTRIBUTION:
  • Controlled Copy No. 01  :          Head Quality Assurance
  • Controlled Copy No. 02  :         Head Quality Control
  • Master Copy                    :          Quality Assurance Department
  • ABBREVIATIONS:
No.:Number
SOP:Standard Operating Procedure
QC:Quality Control
  • REVISION HISTORY:             

CHANGE HISTORY LOG

Revision No.Details of ChangesReason for ChangeEffective Date
00New SOPNot ApplicableTo Be Recorded Manual

Annexure-I
CALIBRATION RECORD OF LAB OVEN

Click the link for download word file copy of this document:
https://pharmaguidehub.com/product/calibration-of-laboratory-oven/

Share this article
Shareable URL
Prev Post

CALIBRATION OF DISINTEGRATION TEST APPARATUS

Next Post

CALIBRATION OF UV CABINET

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

HANDLING OF DESICCATOR

To lay down the procedure for handling and Maintenance of Desiccator. This SOP is applicable for the procedure…

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading