HANDLING AND MAINTENANCE OF IMPURITY STANDARD STOCK SOLUTION

  • PROCEDURE:
  • General Instructions
  • Ensure that the MSDS (Material Safety Data Sheet) of the respective official material is available in the laboratory and all the precautions provided in them are followed.
  • Standard Solutions for Qualitative Analysis:
  • This procedure is not applicable for solutions from which quantitative parameters such as “area, absorbance, RSD (Relative Standard Deviation) etc.” are measured.
  • Prepare a standard solution (Reference standard, in-house reference standard, working standard, impurity standard etc.), which are used for qualitative system suitability parameters such as “resolution factor, capacity factor, tailing factor/symmetrical factor, theoretical plates, etc.,” are measured as per relevant test procedure and shall not be maintained the stock verification.
  • Store the standard solutions in tightly closed bottles or in sealed vials at a temperature between 2°C and 8°C, unless otherwise specified. Label these bottles/vials using the label as per Format-III.
  • Prepare and allot the usage card as given in Format-I and fill the details accordingly.
  • The system suitability solution can be used until the area or area ratio difference of the principal peak between initial analysis and current analysis is within ± 20%.
  • The area difference shall be calculated as follows
  • The solution shall be used as long as it is clear, free from turbidity and the solution complies within the acceptance criteria of analytical parameters.
  • Whenever usage of solution, enter the details of observed parameters, as given in Format-II.
  • The usage is valid only if the observed parameter values are within the acceptance criteria. Else discard the solution.
  • Destruction procedure of solutions: Deface the label and send the solutions to SHE dept, destruction shall be followed as per the SOP.
  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-IStock Solution Usage Card
Annexure-IISuitability details of the solution
Annexure-IIIStock solution Label

Annexure-I
STOCK SOLUTION USAGE CARD

STOCK SOLUTION USAGE CARD

Annexure-II
SUITABILITY DETAILS OF THE SOLUTION

SUITABILITY DETAILS OF THE SOLUTION

Annexure-III
STOCK SOLUTION LABEL

STOCK SOLUTION LABEL
Share this article
Shareable URL
Prev Post

CALIBRATION OF SHIMADZU HPLC WITH EMPOWER SOFTWARE (MODEL LC-2010CHT)

Next Post

HPLC CHROMATOGRAPHIC ANALYSIS AND DOCUMENTATION

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

HANDLING OF GLASSWARE

To lay down the procedure for handling of glassware. This SOP is applicable for the procedure for handling of…

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading