CLEANING OF AIR SAMPLER (Make: MERCK)

  • PROCEDURE:
  • CLEANING PROCEDURE:
  • Air sampler housing:
  • The external surface of the air sampler should be cleaned with commercial disinfectant solution or 70 % isopropanol/water solution).
  • Spray the cleaning solution directly on the instrument then wipe with a sterile towel or spray the towel, then wipe. All cleaning/disinfectant solution should be removed/wiped dry properly before using the air sampler.
  • Sampling head:
  • The perforated lid and dust cover can be autoclaved for 20 minutes at 121 °C. In case of dirt/blocked holes of the perforated lid: clean them with a needle, or by using an ultrasonic bath.
  • Note: Always cover the perforated lid with the dust cover when the air sampler is not in use.
  • Note: The heads should not be allowed to hit on each other.