- PROCEDURE FOR THE FUMIGATION IN MICRO LAB:
- Fumigation in a micro lab is a process of using a gas or vapor to decontaminate the lab environment, eliminating microbial growth, spores, and other pathogens. It’s a powerful method for sterilization.
- The microbiology testing area comprises LAF Room, Cool Zone and Change rooms.
- Ensure that the LAF rooms shall be free from all the used utensils, culture media plates and Petri plates to be incubated.
- Switch “OFF” the Laminar air flow units.
- Switch “OFF” the Air handling Unit.
- Ensure the safety precautions while doing the fumigation activity.
- Weigh about 5.0 gm of potassium Permanganate and place in a non-reactive wide mouth plastic container at each location.
- Perform the fumigation as per the flow direction mentioned below:
- LAF room
- Change room-2
- Cool zone
- Change room-1
Note: Prior to fumigation activity, wear safety goggles, nose mask and gloves.
- Subsequently, add about 35 ml of formaldehyde solution to the container containing potassium permanganate.
- Immediately come out of the respective room.Record the fumigation details in Format-I.
- Fumigation by using Fogger machine:
- Prepare the disinfectant solution for fumigation as indicated by manufacturer.
- Pour the prepared liquid through the funnel on the lid, use hand gloves, goggles and mask.
- Position the fogger machine in a corner diagonally on stool or table in an angle of 45.
- Plug in the Fogger machine to ensure gr connection.
CAUTION: Always ensure “Air filter” is in place before switching on the fogger. Replace filter every 4 months
- Set the timer to desired minutes of fogging. For 1 lit = 20-25 min (in fixed flow mode)
- Switch on the fogger machine and leave the clean room and seal it as “Area under Fumigation”.
- Record the fumigation details in Format-II.
CAUTION: Fog out 500 ml of plain water after every use, to clean internal tubing’s never tamper with the nozzle. It may lead to permanent damage.
Note: Ensure all the doors shall be closed.
- After the completion of fumigation, enter the area with safety goggle, nose mask and gloves.
- Remove the Plastic containers from the area.
- If any spillage is observed, handle the spillage as per SOP.
- Switch “ON” the Air Handling Unit and put the AHU under Exhaust mode for about 2 hrs.Clean the Microbiology testing area as per schedule.
- Fumigation shall be carried out once in a week.
- ANNEXURES:
ANNEXURE NO. | TITLE OF ANNEXURE |
Annexure-I | Fumigation record |
Annexure-II | Fumigation by fogger machine record |
- ABBREVIATIONS:
No. | : | Number |
hrs. | : | hours |
LAF | : | Laminar Air Flow |
AHU | : | Air Handling unit |
SOP | : | Standard Operating Procedure |
QC | : | Quality Control |
Annexure-I
FUMIGATION RECORD
Annexure-II
FUMIGATION BY FOGGER MACHINE RECORD