PROCEDURE FOR CALIBRATION OF POLARIMETER

  • PROCEDURE:
  • Make – Rudolph Research Analytical
  • Operate instrument as per operation SOP.
  • Sample compartment shall be kept empty while switching on instrument.
  • Switch on the instrument by using the ON/OFF switch present on backside of the instrument.
  • Keep the instrument ON condition at least for10 minutes before operation and wait until the reading on the display come to stable position.
  • Press the Zero key and wait until the display shows 0.00 reading.
  • Temp Trol Quartz control plate shall be placed right side in the trough.
  • Insert the built in temperature probe into the Temp Trol Quartz control plate.
  • Set the temperature to 20°C and allow it to stabilize.
  • Measure the optical rotation of the Quartz control plate at wavelength mentioned below.
Optical rotation of the Quartz control

Note: To the above acceptance limits add instrument Accuracy ± 0.01.

  • Temperature Calibration:
  • Select temperature calibration by pressing the temperature calibration key.
  • Place the temperature validation cell into the trough on to the right hand side of the trough.
  • Be sure the foot of Temp trol cell is clean and sitting flat against the clean temp trol temperature transmission surface to allow for maximum heating and cooling efficiency.
  • The built in temperature probe shall be placed in the temperature probe well of the temperature validation cell and tighten the finger nut to a secure.
  • Fill the temperature validation cell with water by adding the water into the temperature validation cell.
  • Fill the cell until the water level is just above the copper sleeve inside the thermometer well.
  • Insert a calibrated thermometer into the thermometer well of the temperature validation cell.
  • Set the temperature to 20°C and allow the built in probe temperature reading to stabilize at 20°C ± 0.1°C for 10 minutes and note down the temperature of the calibrated thermometer.
  • The same procedure shall be followed for 25°C and 30°C respectively and note down the reading.
  • Acceptance criteria shall be ± 0.2°C
  • Enter the details of the usage in the User logbook.
  • ANNEXURES: 
ANNEXURE NO.TITLE OF ANNEXURE
Annexure-ICalibration record of Polari meter

Annexure-I
CALIBRATION RECORD OF POLARIMETER

CALIBRATION RECORD OF POLARIMETER
Share this article
Shareable URL
Prev Post

SAMPLING OF RAW MATERIAL

Next Post

REGISTRATION OF SAMPLES IN QUALITY CONTROL

Leave a Reply

Your email address will not be published. Required fields are marked *

Read next

OPERATION OF VACUUM PUMP

To lay down the procedure for operation of vacuum pump. This SOP is applicable to the procedure for Operation of…

Discover more from PharmaGuideHub

Subscribe now to keep reading and get access to the full archive.

Continue reading